How True Trl/Lrm Works (Option 400 Only); The Trl Calibration Procedure - Agilent Technologies 8719ES User Manual

Network analyzer
Hide thumbs Also See for 8719ES:
Table of Contents

Advertisement

Operating Concepts
TRL*/LRM* Calibration (ES Models Only)
Transmission magnitude uncertainty = E
where:
E
= effective directivity
D
E
= effective reflection tracking
R
E
= effective source match
S
E
= effective load match
L
E
= effective crosstalk
X
E
= effective transmission tracking
T
S
= S-parameters of the device under test
xx

How True TRL/LRM Works (Option 400 Only)

The TRL implementation with Option 400 requires a total of fourteen measurements to
quantify ten unknowns as opposed to only a total of twelve measurements for TRL*. (Both
include the two isolation error terms.)
Because of the four-sampler receiver architecture of Option 400, additional correction of
the source match and load match terms is achieved by measuring the ratio of the two
"reference" receivers during the thru and line steps. These measurements characterize the
impedance of the switch and associated hardware in both the forward and reverse
measurement configurations. They are then used to modify the corresponding source and
load match terms (for both forward and reverse).
The Option 400 configuration with TRL established a higher performance calibration
method over TRL* when making in-fixture measurements, because all significant error
terms are systematically reduced. With TRL*, the source and load match terms are
essentially that of the raw, "uncorrected" performance of the hardware.

The TRL Calibration Procedure

Requirements for TRL Standards
When building a set of TRL standards for a microstrip or fixture environment, the
requirements for each of these standard types must be satisfied.
Types
Requirements
THRU
(Zero length)
• No loss. Characteristic impedance (Z
• S
21
• S
11
THRU
(Non-zero
length)
• Z
0
the average impedance is used.)
7-72
X
= 1 ∠0°
= S
12
= S
= 0
22
of the thru must be the same as the line. (If they are not the same,
+ E
S
+ E
S
S
+ E
T
21
S
11
21
) need not be known.
0
S
S
L
22
21

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

8719et8720es8720et8722es8722et

Table of Contents