Eliminating Unwanted Mixing And Leakage Signals - Agilent Technologies 8719ES User Manual

Network analyzer
Hide thumbs Also See for 8719ES:
Table of Contents

Advertisement

Making Mixer Measurements (Option 089 Only)
Measurement Considerations

Eliminating Unwanted Mixing and Leakage Signals

By placing filters between the mixer's IF port and the receiver's input port, you can
eliminate unwanted mixing and leakage signals from entering the analyzer's receiver.
Filtering is required in both fixed and broadband measurements. Therefore, when
configuring broad-band (swept) measurements, you may need to trade some measurement
bandwidth for the ability to more selectively filter signals entering the analyzer receiver.
Figure 2-4 Example of Conversion Loss versus Output Frequency without
Correct IF Signal Path Filtering
2-6