The "format" choice is selected by the current selection under the FORMAT menu. To
select the DB format, the FORMAT must be LOG MAG. For MA, the FORMAT must be
LIN MAG (unlike CITIfile), and all other FORMAT selections will output RI data.
The S2P data will always represent the format array data, including effects of electrical
delay and port extensions. A CITIfile will be saved at the same time. To be consistent with
previous versions, the CITIfile data saved will represent the DATA array (corrected data)
without effects of electrical delay or port extensions.
CAUTION
Using the smoothing feature or saving data displayed in time domain format
may result in invalid S2P data. Avoid using these functions when saving S2P
files.
Here is an S2P example file for an 11 point measurement of a 20 dB attenuator:
#
HZ
S
DB
R 50
! Network Analyzer HP8753E.0611
−56.74
50000000
−53.015
250000000
−52.094
450000000
−51.758
650000000
−50.95
850000000
−50.235
1050000000
−49.883
1250000000
−48.477
1450000000
−48.462
1650000000
−47.503
1850000000
−46.938
2050000000
−20.219
−2.0132
15.178
−20.373
−10.241
1.7331
−20.391
−18.555
5.8173
−20.189
−26.18
8.02
−20.163
−34.743
11.472
−20.178
−42.682
9.3562
−20.142
−50.854
9.2574
−20.201
−58.917
5.9944
−20.161
−67.008
3.5156
−20.15
−74.862
1840
−5.6538
−20.167
−83.048
Printing, Plotting, and Saving Measurement Results
Saving Measurement Results
−20.15
−1.6658
−36.188
−20.377
−10.029
−33.974
−20.387
−17.96
−31.287
−20.112
−26.061
−29.427
−20.198
−34.195
−24.719
−20.19
−42.289
−25.102
−20.223
−50.407
−27.582
−20.21
−58.436
−33.828
−20.188
−66.587
−44.184
−20.208
−74.616
−36.893
−20.256
−82.874
−30.385
−123.52
−40.215
61.778
153.37
−137.83
−81.096
−25.509
35.237
62.912
35.384
74.001
4- 43