Making Measurements
Using Test Sequencing to Test a Device
To create a second sequence that will perform a desired measurement function,
decrement the loop counter, and call itself until the loop counter value is equal to
zero, press:
Seq
NEW SEQ/MODIFY SEQ
Trans: FWD S21 (B/R)
Meas
Scale Ref
Marker Search
SPECIAL FUNCTIONS
Seq
DECR LOOP COUNTER
SEQUENCE 2 SEQ2
DONE SEQ MODIFY
Seq
This will create a displayed list as shown:
SEQUENCE LOOP 2
Start of Sequence
Trans:FWD S21 (B/R)
SCALE/DIV
AUTO SCALE
MKR Fctn
SEARCH MAX
DECR LOOP COUNTER
IF LOOP COUNTER <> 0 THEN DO
SEQUENCE 2
To run the loop sequence, press:
Preset
SEQUENCE 1 SEQ1
Generating Files in a Loop Counter Example Sequence
This example shows how to increment the names of files that are generated by a sequence
with a loop structure.
NEW SEQ/MODIFY SEQ
Seq
SPECIAL FUNCTIONS
LOOP COUNTER 7
SELECT DISK
Save/Recall
RETURN
DEFINE DISK-SAVE
SET ADDRESSES
Local
DO SEQUENCE
Seq
DONE SEQ MODIFY
NEW SEQ/MODIFY SEQ
1-116
or on ET models:
AUTO SCALE
SEARCH: MAX
DECISION MAKING
IF LOOP COUNTER 0
SEQUENCE 1 SEQ 1
DECISION MAKING
x1
INTERNAL DISK
DATA ONLY ON
PLOTTER PORT
SEQUENCE 2
SEQUENCE 2 SEQ 2
SEQUENCE 2 SEQ2
TRANSMISSN
DISK