Determine The Electrical Delay - Agilent Technologies 8719ES User Manual

Network analyzer
Hide thumbs Also See for 8719ES:
Table of Contents

Advertisement

Determine the Electrical Delay

This procedure determines the electrical delay of adapter A3 using a short.
1. Refer to
Figure 6-29
2. Perform a 1-port calibration at "Reference Port 1". Refer to Step A of
This 1-port calibration can either be a manual calibration or an ECal.
Figure 6-29 Determining the Electrical Delay Setup
3. Connect the A3 adapter to "Reference Port 1" as shown in Step B of
Connect a short to the open end of the A3 adapter.
4. Measure the delay of the adapter by pressing
5. Divide the resulting delay measurement by 2 to determine the delay of the thru and the
short in one direction.
6. Determine the offset delay of the calibration short by examining the define standard
menu (see
"Define Standard Menus" on page
7. Subtract the offset delay of the short (determined in step 6) from the delay of the thru
and the short in one direction (determined in step 5). The result is electrical delay of the
thru. This value is used in the next procedure.
while performing the steps in this procedure.
Calibrating for Increased Measurement Accuracy
Adapter Removal Using ECal (ES Analyzers Only)
.
Format
DELAY
7-59).
Figure
6-29.
Figure
6-29.
6- 73

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

8719et8720es8720et8722es8722et

Table of Contents