Creating Single Point Limits - Agilent Technologies 8719ES User Manual

Network analyzer
Hide thumbs Also See for 8719ES:
Table of Contents

Advertisement

Figure 1-61 Sloping Limit Lines

Creating Single Point Limits

In this example procedure, the following limits are set:
• from −23 dB to −28.5 dB at 141 MHz
• from −23 dB to −28.5 dB at 126.5 MHz
1. To access the limits menu and activate the limit lines, press:
LIMIT MENU
System
CLEAR LIST
YES
2. To designate a single point limit line, as shown in
pointers:
• downward pointing, indicating the upper test limit
• upward pointing, indicating the lower test limit
Press:
ADD
STIMULUS VALUE
UPPER LIMIT
LIMIT TYPE
SINGLE POINT
ADD
STIMULUS VALUE
UPPER LIMIT
LOWER LIMIT
LIMIT TYPE
SINGLE POINT
LIMIT LINE
141
M/µ
−23
LOWER LIMIT
x1
RETURN
126.5
M/µ
−23
x1
−28.5
DONE
x1
RETURN
Using Limit Lines to Test a Device
LIMIT LINE ON
EDIT LIMIT LINE
Figure
1-62, you must define two
−28.5
DONE
x1
Making Measurements
1- 77