Making Measurements
Using Test Sequencing
The previous keystrokes will create a displayed list as shown:
Start of Sequence
RECALL PRST STATE
Trans: FWD S21 (B/R)
LOG MAG
CENTER
134 M/u
SPAN
50 M/u
SCALE/DIV
AUTO SCALE
4. To complete the sequence creation, press:
DONE SEQ MODIFY
Seq
When you create a sequence, the analyzer stores it in volatile memory where
CAUTION
it will be lost if you switch off the instrument power (except for sequence #6
which is stored in the analyzer non-volatile memory). However, you may store
sequences to a floppy disk.

Running a Sequence

To run a stored test sequence, press:
and the softkey labeled with desired sequence number.
Preset
or, press:
DO SEQUENCE
Seq

Stopping a Sequence

To stop a sequence before it has finished, press

Editing a Sequence

Deleting Commands
1. To enter the creation/editing mode, press:
NEW SEQ/MODIFY SEQ
Seq
2. To select the particular test sequence you wish to modify (sequence 1 in this example),
press:
SEQUENCE 1 SEQ1
1-100
and the softkey labeled with the desired sequence number
.
Local