Assign the Standards to the Various LRM Classes
8. To assign the calibration standards to the various TRL calibration classes, press:
Cal
CAL KIT
9. Since you previously designated standard #1 for the REFLECT standard, press:
1
x1
10.Since you previously designated standard #3 for the LINE/MATCH standard, press:
TRL LINE OR MATCH
11.Since you previously designated standard #4 for the THRU/LINE standard, press:
4
TRL THRU
12.To complete the specification of class assignments, press:
SPECIFY CLASS DONE
Label the Classes
NOTE
To enter the following label titles, an external keyboard may be used for
convenience.
13.Press
LABEL CLASS
14.Change the label of the "TRL REFLECT" class to "LRMSHORT."
15.Change the label of the "TRL LINE OR MATCH" class to "LRMLOAD."
16.Change the label of the "TRL THRU" class to "LRMTHRU."
17.Press
LABEL CLASS DONE
Label the Calibration Kit
18.Press
LABEL KIT
"LRM KIT1" and press
19.To save the newly defined kit into nonvolatile memory, press:
KIT DONE (MODIFIED)
NOTE
Refer to
information about saving modified calibration kits, along with calibration
data and instrument states, to a disk.
MODIFY
SPECIFY CLASS MORE
3
x1
x1
MORE
MORE
.
and create a label up to 8 characters long. For this example, enter
.
DONE
SAVE USER KIT
"Saving Modified Calibration Kits to a Disk" on page 7-65
Calibrating for Increased Measurement Accuracy
MORE
.
LRM Error Correction
TRL REFLECT
for
6- 55