Making Measurements
Using Limit Lines to Test a Device
Figure 1-62 Example Single Points Limit Line
Editing Limit Segments
This example shows you how to edit the upper limit of a limit line.
1. To access the limits menu and activate the limit lines, press:
LIMIT MENU
System
2. To move the pointer symbol (>) on the analyzer display to the segment you wish to
modify, press:
SEGMENT
followed by
.
x1
3. To change the upper limit (for example, −20) of a limit line, press:
EDIT
UPPER LIMIT
Deleting Limit Segments
1. To access the limits menu and activate the limit lines, press:
LIMIT MENU
System
2. To move the pointer symbol (>) on the analyzer display to the segment you wish to
delete, press:
SEGMENT
followed by
.
x1
3. To delete the segment that you have selected with the pointer symbol, press:
DELETE
1-78
LIMIT LINE
or
repeatedly OR
−20
DONE
x1
LIMIT LINE
or
repeatedly OR
LIMIT LINE ON
EDIT LIMIT LINE
and enter the segment number
SEGMENT
LIMIT LINE ON
EDIT LIMIT LINE
and enter the segment number
SEGMENT