Figure 2-16 Connections for Power Meter Calibration
3. Select the analyzer as the system controller:
Local
SYSTEM CONTROLLER
4. Set the power meter's address:
SET ADDRESSES
ADDRESS: P MTR/GPIB
5. Select the appropriate power meter by pressing
model number is displayed (436A or 438A/437).
NOTE
The E4418B and E4419B power meters have a "437 emulation" mode. This
allows these power meters, with an 848X-series power sensor, to be used with
the network analyzer. In this step, when selecting a power meter, choose the
438A/437 selection.
6. Press
Cal
PWRMTR CAL
enter the correction factors as listed on the power sensor. Press
(where fff is the frequency in MHz)
fff
calibration factor number)
DONE
RETURN
7. Perform a one sweep power meter calibration over the IF frequency range at −5 dBm
(−10 dBm, 8722ES):
ONE SWEEP
−5
−10
(
x1
TAKE CAL SWEEP
(where aa is the power meter GPIB address)
aa
LOSS/SENSR LISTS
DONE
x1
.
, 8722ES)
x1
Making Mixer Measurements (Option 089 Only)
High Dynamic Range Swept RF/IF Conversion Loss
POWER MTR [ ]
CAL FACTOR
CAL FACTOR
M/µ
for each correction factor. When finished, press
x1
until the correct
and
SENSOR A
ADD
FREQUENCY
(where nnn is the
nnn
2- 21