6. To select the correction type, press
and select the correction type.
If you want to make measurements in the forward direction, press:
S11/S21 ENH. RESP.
For ES analyzers, if you want to make measurements in the reverse direction, press:
S22/S12 ENH. RESP.
7. Connect a shielded open circuit to PORT 1 (or PORT 2 for reverse measurements).
Include any adapters that you will have in the device measurement. That is,
NOTE
connect the standard to the particular connector where you will connect your
device under test.
Figure 6-7 Standard Connections for Enhanced Response Calibration
8. To measure the standard, when the displayed trace has settled, press:
OPEN
The analyzer displays WAIT - MEASURING CAL STANDARD during the standard
measurement. The analyzer underlines the
standard.
9. Disconnect the open, and connect a short circuit to the test port.
10.To measure the device, when the displayed trace has settled, press:
SHORT
The analyzer measures the short circuit and underlines the
11.Disconnect the short, and connect an impedance-matched load to the test port.
CALIBRATE MENU
or on ET models:
TRAN/REFL ENH. RESP.
OPEN
Calibrating for Increased Measurement Accuracy
Enhanced Frequency Response Error Correction
ENHANCED RESPONSE
softkey after it measures the
softkey.
SHORT
6- 23