Figure 6-2 Standard Connections for a Response Error Correction for
Reflection Measurement
7. To measure the standard when the displayed trace has settled, press
.
OPEN
If the calibration kit you selected has a choice between male and female calibration
standards, remember to select the sex that applies to the test port and not the standard.
The analyzer displays WAIT - MEASURING CAL STANDARD during the standard
measurement. The analyzer underlines the softkey that you selected after it finishes
the measurement, and computes the error coefficients.
This calibration allows only one standard to be measured. If you press the
NOTE
wrong key for a standard, press
standard. Do not use a thru standard for a reflection response correction.
NOTE
You can save or store the measurement correction to use for later
measurements, that use the same measurement parameters. Refer to the
Chapter 4 , "Printing, Plotting, and Saving Measurement Results"
procedures.
This completes the response correction for reflection measurements. You can connect
and measure your device under test.
Calibrating for Increased Measurement Accuracy
Frequency Response Error Corrections
again and choose the correct
RESPONSE
or
SHORT
for
6- 13