Characterizing Microwave Systematic Errors - Agilent Technologies 8719ES User Manual

Network analyzer
Hide thumbs Also See for 8719ES:
Table of Contents

Advertisement

Operating Concepts
Measurement Calibration
Frequency Response (Tracking)
This is the vector sum of all test setup variations in which magnitude and phase change as
a function of frequency. This includes variations contributed by signal-separation devices,
test cables, adapters, and variations between the reference and test signal paths. This
error is a factor in both transmission and reflection measurements.
For further explanation of systematic error terms and the way they are combined and
represented graphically in error models, refer to the
"Characterizing Microwave
Systematic Errors" on page
7-41.

Characterizing Microwave Systematic Errors

One-Port Error Model
In a measurement of the reflection coefficient (magnitude and phase) of a test device, the
measured data differs from the actual, no matter how carefully the measurement is made.
Directivity, source match, and reflection signal path frequency response (tracking) are the
major sources of error. See
Figure
7-24.
Figure 7-24 Sources of Error in a Reflection Measurement
7- 41