The preset state of the instrument can be configured so that interpolated
NOTE
error correction is on or off. Press
USER SETTINGS
the preset state of interpolated error correction.
System performance is unspecified when using interpolated error correction. The quality of
the interpolated error correction is dependent on the amount of phase shift and the
amplitude change between measurement points. If phase shift is no greater than 180° per
approximately five measurement points, interpolated error correction offers a great
improvement over uncorrected measurements. The accuracy of interpolated error
correction improves as the phase shift and amplitude change between adjacent points
decrease. When you use the analyzer in linear frequency sweep, perform the original
calibration with at least 30 points per 1 GHz of frequency span for greatest accuracy with
interpolated error correction.
Interpolated error correction is available in three sweep modes: linear frequency, power
sweep, and CW time.
NOTE
If there is a valid correction array for a linear frequency sweep, this may be
interpolated to provide correction at the CW frequency used in power sweep
or CW time modes. This correction is part of the interpolated error-correction
feature.

Error-Correction Stimulus State

Error correction is only valid for a specific stimulus state, which you must select before you
start a correction. If you change any of the following parameters, you will invalidate the
correction and the analyzer will switch the correction off (unless the interpolated error
correction feature is activated):
• frequency range
• number of points
• sweep type
The error correction quality may be degraded (Cor changes to C∆), if you change the
following stimulus state parameters:
• sweep time
• system bandwidth
• output power
If correction is turned off or shows C∆, the original stimulus state can be recalled by, first,
turning interpolation off (
(
CORRECTION ON off
PRESET SETTINGS
INTERPOL ON off
).
Calibrating for Increased Measurement Accuracy
CONFIGURE MENU
System
CAL INTERP ON off
), and toggling correction off and then on
Calibration Considerations
to configure
6- 9