Operating Concepts
Measurement Calibration
Figure 7-32 Open Circuit Termination
This completes the calibration procedure for one port devices.
Device Measurement
Now the unknown is measured to obtain a value for the measured response, S
, at each
11M
frequency. Refer to
Figure
7-33.
Figure 7-33 Measured S
11
This is the one-port error model equation solved for S
. Since the three errors and S
11A
11M
are now known for each test frequency, S
can be computed as follows:
11A
(
)
S
E
11M
DF
S
=
---------------------------------------------------------------- -
11A
(
)
+
E
E
S
E
RF
SF
11M
DF
For reflection measurements on two-port devices, the same technique can be applied, but
the test device output port must be terminated in the system characteristic impedance.
This termination should have as low a reflection coefficient as the load used to determine
directivity. The additional reflection error caused by an improper termination at the test
device's output port is not incorporated into the one-port error model.