Using Continuous Correction Mode - Agilent Technologies 8719ES User Manual

Network analyzer
Hide thumbs Also See for 8719ES:
Table of Contents

Advertisement

Calibrating for Increased Measurement Accuracy
Power Meter Measurement Calibration

Using Continuous Correction Mode

You can set the analyzer to update the correction table at each sweep (as in a leveling
application), using the continuous sample mode. When the analyzer is in this mode, it
continuously checks power at every point in each sweep. You must keep the power meter
connected as shown in
the speed is limited by the power meter.
NOTE
You may level at the input of a device under test, using a 2-resistor power
splitter or a directional coupler before the device; or level at the output of the
device using a 3-resistor power splitter or a bidirectional coupler after the
device.
Figure 6-11 Continuous Correction Mode for Power Meter Calibration
1. Connect a power splitter or directional coupler to the port supplying RF power to your
test device, as shown in
2. Set test port power to approximate desired leveled power.
3. Press
Cal
PWRMTR CAL
analyzer to maintain at the input to your test device. Compensate for the power loss of
the power splitter or directional coupler in the setup.
4. If you want the analyzer to make more than one power measurement at each frequency
data point, press
iterations).
If you increase the number of readings, the power meter correction time will
substantially increase.
5. Press
PWRMTR CAL
Cal
power meter correction.
6-38
Figure
6-11. This mode is also known as power meter leveling, and
Figure
6-11.
and enter the test port power level that you want the
NUMBER OF READINGS
EACH SWEEP
(where n = the number of desired
n
x1
TAKE CAL SWEEP
to activate the