INSTRUMENT MODE FREQ OFFS MENU
System
LO FREQUENCY 800
4. To set the analyzer to the desired power sweep range, press:
Power
PWR RANGE MAN
−10
Start
Stop
10
x1
5. Make the connections, as shown in
CAUTION
To prevent connector damage, use an adapter (part number 1250-1462) as a
connector saver for R CHANNEL IN.
Figure 2-28 Connections for the First Portion of Conversion Compression
Measurement
6. To view the absolute input power to the analyzer's R channel, press:
INPUT PORTS
Meas
7. To store a trace of the receiver power versus the source power into memory and view
data/memory, press:
Display
DATA→MEM
DATA/MEM
This removes the loss between the output of the mixer and the input to the receiver, and
provides a linear power sweep for use in subsequent measurements.
M/µ
POWER RANGES RANGE 0
x1
Figure
R
Making Mixer Measurements (Option 089 Only)
Conversion Compression Using the Frequency Offset Mode
2-28.
2- 41