Making Measurements

Making a Basic Measurement

Making a Basic Measurement

There are five basic steps when you are making a measurement.
1. Connect the device under test and any required test equipment.
CAUTION
Damage may result to the device under test (DUT) if it is sensitive to the
analyzer's default output power level. To avoid damaging a sensitive DUT, be
sure to lower the output power before connecting the DUT to the analyzer.
2. Choose the measurement parameters.
3. Perform and apply the appropriate error-correction.
4. Measure the device under test (DUT).
5. Output the measurement results.
This example procedure shows you how to measure the transmission response of a
bandpass filter.
Step 1. Connect the device under test and any required test
equipment.
Make the connections as shown in
Figure 1-1 Basic Measurement Setup

Step 2. Choose the measurement parameters.

Press
.
Preset
To set preset the analyzer to the "Factory Preset" conditions, press the
PRESET: FACTORY
1-4
Figure
1-1.
softkey if it is not selected. Then press
.
Preset