Making Measurements
Using Markers
Figure 1-29 Example of Searching for the Minimum Amplitude Using a Marker
Searching for a Target Amplitude
1. Press
Marker Search
2. Press
SEARCH: TARGET
measurement trace.
3. If you want to change the target amplitude value (default is −3 dB), press
and enter the new value from the front panel keypad. You may also press
TARGET VALUE
Marker Search
4. If you want to search for multiple responses at the target amplitude value, press
and
SEARCH LEFT
Figure 1-30 Example of Searching for a Target Amplitude Using a Marker
1-40
to access the marker search menu.
to move the active marker to the target point on the
to enter the new value.
.
SEARCH RIGHT
TARGET