Chapter 7 Periodic Maintenance
This completes the greasing procedure for the justifier motor unit assembly gear mechanism unit.
Go to
"7.2.11 Attachment of Exterior Panels and Insertion Unit (Page 7-14)"
exterior panels and insertion slot unit.

7.2.11 Attachment of Exterior Panels and Insertion Unit

Attach the removed exterior panels and insertion slot unit after completion of the maintenance.
MDU
First Front Back Side
7-14
1
First Front Side
2
CN1
CN2
3
First Front Back Side
and attach the removed
Attach the insertion unit to the equipment.
5 hex/Phillips-head screws (M4 x 8)
Take care of the cable routing and not to
pinch the cable.
Caution
Connect the cables on the insertion unit to the
MDU.
MCN1
CN1 (MDU)
MCN2
CN2 (MDU)
Attach the circuit board cover.
3 screws (M3 x 6)