Second Front Back Side
8
See
"
Installation Procedures (Page 5-8)"
and Insertion Unit"
to install the second front back panel.
3
Measure the wire tension.
Strike the wire around the center of the belt using ball
of a finger, and measure the vibration noise toward the
edge of the wire.
Caution
See
"8.5 Operation of the Acoustic Wave Tension Meter
(Page 8-13)"
meter.
Specified value is following.Perform the adjustment
according to
Specified
value
Confirmation
method
4
Loosen the fixing screws of the wire holding
assembly.
1 hex/Phillips-head screw (M4 x 12)
5
Adjust by turning the adjustment screw so the
Adjustable Screw
tension is within the specified value.
1 screw (M4 x 12)
Setscrew
6
Fix the wire holding assembly by tightening the
fixing screw.
1 hex/Phillips-head screw (M4 x 12)
7
Release the simple fixing lock.
Lock will be released by pulling up the blue knob.
Important
in
"5.2.3 Removing/Installing the Exterior Panel
Chapter 6 Adjustment
Do not strike the wire with your fingernail.
for how to use the acoustic wave tension
Step 4
-
Step 6
if it is out of range.
Tension: 30 ± 5 N
Acoustic wave tension meter
Input data
Weight (MASS): 1.4 g/m
Width: 1.0 mm
Span: 555 mm
Push the optical unit slightly towards the
back when pulling up the knob. Fixing shaft
might move down, and get locked again if
the optical unit is at its hithermost position.
Also, move the optical unit to confirm that
the fixing shaft is not lowered after
unlocking.