Konica Minolta REGIUS 110 Service Manual page 231

Direct digitizer
Table of Contents

Advertisement

First Front Side
10
See
"
Installation Procedures (Page 5-8)"
and Insertion Unit"
to install the first front panel.
Now, you have finished with the procedures to replace the transformer.
Chapter 5 Disassembly and Assembly
4
Remove the transformer unit assembly.
2 hex/Phillips-head screws (M4 x 8)
5
Remove the transformer.
4 screws (M4 x 6)
6
Install the new transformer.
4 screws (M4 x 6)
Press it in a direction of the arrow shown in the figure
to fix it.
7
See
Step 4
to install the transformer unit
assembly.
2 hex/Phillips-head screws (M4 x 8)
8
Connect the connectors removed in
fix them using the wiring band and clamps.
Connectors at 10 locations
2 cables to the interlock relay 1
9
Install the electric component mount panel that
was removed in
Step
2.
8 hex/Phillips-head screws (M4 x 8)
in
"5.2.3 Removing/Installing the Exterior Panel
Step 3
and
5-111