Chapter 7 Periodic Maintenance
Cleaning cassette
absorption magnet
Greasing LM guide
Greasing transporter lead
screw unit
Greasing transporter motor
unit assembly gear
mechanism
Greasing of release shaft
holder
Greasing justifier motor unit
assembly gear mechanism
Attachment of exterior
panels and the insertion
unit
Move to the installed
location
Power ON
Maintenance finished
7-4
Item
Clean the surface of the magnet on the cassette absorption plate.
See
"7.2.6 Cleaning Cassette Absorption Magnet (Page 7-10)"
description.
Grease the LM guide using a grease for metal.
See
"7.2.7 Greasing LM Guide (Page 7-11)"
Apply grease to the transporter lead screw unit and the transporter motor
unit assembly gear mechanism unit.
See
"7.2.8 Greasing Transporter Lead Screw Unit and Transporter Motor
Unit Assembly Gear Mechanism Unit (Page 7-12)"
Apply grease to the release shaft holder.
See
"7.2.9 Greasing of Release Shaft Holder (Page 7-13)"
description.
Apply grease to the justifier motor unit assembly gear mechanism unit.
See
"7.2.10 Greasing Justifier Motor Unit Assembly Gear Mechanism
(Page 7-13)"
Attach the removed exterior panels and insertion slot unit.
See
"7.2.11 Attachment of Exterior Panels and Insertion Unit (Page 7-
14)"
for description.
Move the equipment back to the installed location if it was moved.
See
"7.2.12 Move to the Installed Location (Page 7-16)"
Connect the Ethernet cable and power cable and turn ON the power.
See
"7.2.13 Power ON (Page 7-16)"