Setscrew
8
See
"
Installation Procedures (Page 5-8)"
and Insertion Unit"
to install the first front panel.
3
Measure the belt tension.
Strike around the center of the belt, and measure the
vibration noise toward the end of the belt.
See
"8.5 Operation of the Acoustic Wave Tension
Meter (Page 8-13)"
tension meter.
Specified value is following. Perform the adjustment
according to
Specified
value
Confirmation
method
4
Loosen the fixing screws of the receiver idler
pulley unit assembly.
3 hex/Phillips-head screws (M4 x 8)
Adjustable
5
Screw
Adjust by turning the adjustment screw so the
tension is within the specified value.
1hexagon socket head bolt (M4 x 20)
1 hexagon nut
Setscrew
6
Fix the receiver idler pulley unit assembly by
tightening the fixing screws.
3 hex/Phillips-head screws (M4 x 8)
7
Attach the electric component mount panel that
was removed in
8 hex/Phillips-head screws (M4 x 8)
in
"5.2.3 Removing/Installing the Exterior Panel
Chapter 6 Adjustment
for how to use the acoustic wave
Step 4
-
Step 6
if it is out of range.
Tension: 44 ± 4 N
Acoustic wave tension meter
Input data
Weight (MASS): 2.5g/m
Width: 9.0mm
Span: 507mm
Step
2.