Konica Minolta REGIUS 110 Service Manual page 68

Direct digitizer
Table of Contents

Advertisement

Chapter 3 Troubleshooting
Error code
S
E
06151 06166 06181 Subscan motor control error
06152 06167 06182 Transporter motor control
06153 06168 06183 Subscan motor control error
References for response method
See
"2.5.4 Subscan Unit (Page 2-14)"
See
"2.5.5 Optical Unit/Eraser Unit (Page 2-15)"
See
"5.5.3 Replacing the Encoder/Wires (Page 5-68)"
See
"5.6.1 Replacing the LMD (Page 5-78)"
See
"5.6.2 Replacing the LMC (Page 5-79)"
3-14
Description of the error
I
(when moving to mechanical
HP)
There was a trouble with the
motor control when the subscan
was moving from the home
position to the mechanical home
position.
error (when moving to read
end position)
There was a trouble with the
motor control when the subscan
was moving from the
mechanical home position to the
read end position.
(when moving to mechanical
home position 2)
There was a trouble with the
motor control when the subscan
was moving from the read end
position to the mechanical home
position.
for the position of the encoder and the subscan motor.
for the procedure to replace the LMD.
for the procedure to replace the LMC.
Response
1. Confirm that the fixing shaft of the
simple fixing lock is not lowered.
2. Confirm there are no foreign objects
on the magnet shaft of the subscan
unit, and remove it if there are any.
3. Confirm that it is installed horizontally.
4. Confirm following and replace the
faulty parts.
Subscan motor (cannot replace
on-site)
Motor wire: CN2 (LMD)
LMD
LMC
for the position of the LMD and the LMC.
for the procedure to replace the encoder.