Konica Minolta REGIUS 110 Service Manual page 181

Direct digitizer
Table of Contents

Advertisement

10
See
"
Installation Procedures (Page 5-8)"
and Insertion Unit"
to install the insertion unit and exterior panel.
Now, you have finished with the procedures to replace the receiver belt.
7
Install the new timing belt.
2 belt presser plates
2 screws for each plate (M3 x 10)
Wiring
guide
Perform installation so that two belt cogs are exposed
for each receiver cog.
8
See
"6.2 Adjust Receiver Belt Tension (Page 6-
4)"
to adjust the belt tension.
9
Install the electric component mount panel that
was removed in
8 hex/Phillips-head screws (M4 x 8)
in
"5.2.3 Removing/Installing the Exterior Panel
Chapter 5 Disassembly and Assembly
Step
3.
5-61