Konica Minolta REGIUS 110 Service Manual page 223

Direct digitizer
Table of Contents

Advertisement

First Front Side
First Front Side
(Brwon)
(Brown)
MDU, etc.
Interlock
Switch
(Black)
(Blue)
Interlock
Digital Power Supply
Relay 1, etc.
7
See
"
Installation Procedures (Page 5-8)"
and Insertion Unit"
to install the first front panel.
Now, you have finished with the procedures to replace the interlock relay 2.
Chapter 5 Disassembly and Assembly
2
Remove the electric component mount panel.
8 hex/Phillips-head screws (M4 x 8)
Loosen all screws.
3
Remove the interlock relay 2.
2 screws (M4 x 6)
4
Unplug the cable and connect it to the new
interlock relay 1.
5
Install the new interlock relay 2.
2 screws (M4 x 6)
6
Install the electric component mount panel that
was removed in
Step
2.
8 hex/Phillips-head screws (M4 x 8)
in
"5.2.3 Removing/Installing the Exterior Panel
5-103

Hide quick links: