Konica Minolta REGIUS 110 Service Manual page 192

Direct digitizer
Table of Contents

Advertisement

Chapter 5 Disassembly and Assembly
Second Front Back Side
31
See
"
Installation Procedures (Page 5-10)"
Door"
to install the second front door.
32
See
"
Installation Procedures (Page 5-14)"
install the optical unit.
Now, you have finished with the procedures to replace the encoder or wires.
5-72
23
Give a single wrapping of the wire around the
encoder.
Follow the descriptions in the figure to wrap.
Important
24
See
Step 12
to fix the other front edge of the wire
using the wire presser plate in the second front
side.
1 screw (M4 x 8)
25
See
Step 10
to slightly tighten the tension in the
wire.
If the tension in the wire is left loosen, the
wire may be disconnected from the encoder
Note
while warming up the wrapping (
26
Move the holding plate back and forth three times
or more to warm up the wire wrapping.
27
See
"6.4 Adjust Wire Tension (Page 6-10)"
adjust the wire tension.
28
Rotate the shaft clockwise and move the push
plate unit slightly toward the first front side
following the instruction in
29
Install the wire cover that was removed in
2 screws (M4 x 8)
Do not drop the wire cover on the wire.
Keep the wire cover away from the
Caution
magnet shaft.
30
Move the push plate unit to the read position.
in
"5.2.4 Removing/Installing the Second Front
in
"5.2.5 Removing/Installing the Optical Unit"
Step 26
).
to
Step
3.
Step
6.
to

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

0902

Table of Contents