Konica Minolta REGIUS 110 Service Manual page 183

Direct digitizer
Table of Contents

Advertisement

5
First Front Side
6
JP50
First Front Side
7
Setscrew
Adjustable screw
Adjustable Screw
Setscrew
8
9
Chapter 5 Disassembly and Assembly
Remove the belt presser plate first, then the
timing belt.
1 tapping screw (M4 x 10)
Incline the justifier guide unit assembly to remove the
screw from the first front side.
Remove the justifier motor connector (JP50).
When removing the connector, hold a notch
on the connector without pulling a cable.
Caution
Giving a strong pull to the cable may cause
it to be broken.
Remove the motor mount board.
1. Loosen the adjustable screws.
2 hexagon socket head bolts (M4 x 20)
2. Remove the setscrews.
3 hex/Phillips-head screws (M4 x 8)
3. Remove the motor mount board.
Remove the gear first, then the timing belt.
Gear: 1 screw (M4 x 8)
Thrust collars: 2 setscrews for each one
Lead the new timing belt and install the gear.
Gear: 1 screw (M4 x 8)
Thrust collars: 2 setscrews for each one
Fix the thrust collar placed far from you, mating the
surfaces of it and the shaft end.Fix the thrust collar
placed toward you, pressing it lightly to the shaft so as
not to have the shaft misaligned after the collar is fixed.
For the upper gear, install it, mating the
surfaces of the thrust collar and the shaft
Note
end for the left side, while pressing lightly for
the right side.
5-63

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

0902

Table of Contents