Chapter 5 Disassembly and Assembly
10
See
"
Installation Procedures (Page 5-8)"
and Insertion Unit"
Now, you have finished with the procedures to replace the Lock Motor.
5-54
4
Remove the cam unit assembly.
5
Remove the Lock Motor.
6
Install the new Lock Motor on the motor mount
board.
7
Install the cam unit assembly.
Fix the gear, aligning the Surface A in the figure.
A
8
Install the lock release motor that was removed in
Step
9
Connect the connector (JP16) that was removed
in
in
to install the first front panel.
2 setscrews
3 screws (M3 x 6)
3 screws (M3 x 6)
2 setscrews
3.
2 screws (M4 x 8)
Step
2.
"5.2.3 Removing/Installing the Exterior Panel

Hide quick links: