Chapter 5 Disassembly and Assembly
5.4.6
Replacing the Receiver HP Sensor
This section describes the instructions on how to replace the receiver HP sensor.
Be sure to turn off the power supply circuit breaker and unplug the power cable from the
equipment whenever you disassemble/assemble the equipment.
Caution
Work outline
Listed below are the number of people and the rough standard of work hours required for performing the
works described in this section.
Personnel
Number
1
Requirements
The works in this section can be performed using only the standard tools.
1
See
"
Removal Procedures (Page 5-6)"
Insertion Unit"
First Front Side
First Front Side
JP43
5-48
Work Hours (steps 2 to 9)
to remove the exterior panel and insertion unit.
in
"5.2.3 Removing/Installing the Exterior Panel and
2
Remove the exterior frame (front).
•
4 screws (M4 x 8)
Loosen all screws.
3
Remove the receiver unit.
•
1 hex/Phillips-head screw (M4 x 8)
4
Unplug the cable from the receiver HP sensor.
•
Connector (JP43)
When removing the connector, hold a notch
on the connector without pulling a cable.
Caution
Giving a strong pull to the cable may cause
it to be broken.
5
Remove the receiver HP sensor.
•
1 screw (M4 x 15)