Konica Minolta REGIUS 110 Service Manual page 154

Direct digitizer
Table of Contents

Advertisement

Chapter 5 Disassembly and Assembly
8
See
"
Installation Procedures (Page 5-8)"
and Insertion Unit"
Now, you have finished with the procedures to replace the shutter close detection sensor.
5-34
3
Remove the bracket from the shutter close
detection sensor.
4
Remove the shutter close detection sensor.
5
Install the new shutter close detection sensor on
the bracket.
6
Install the bracket that was removed in
7
Connect the cables that were unplugged in
2
in
to install the insertion unit and exterior panel.
1 screw (M4 x 8)
1 screw (M4 x 8)
to the shutter close detection sensor.
Connector (JP36)
"5.2.3 Removing/Installing the Exterior Panel
Step
3.
Step