Konica Minolta REGIUS 110 Service Manual page 159

Direct digitizer
Table of Contents

Advertisement

Grounding Cable
JP20
8
See
"
Installation Procedures (Page 5-8)"
and Insertion Unit"
to install the insertion unit and exterior panel.
Now, you have finished with the procedures to replace the shutter unit.
7
Connect the cables that were unplugged in
2
to the shutter unit assembly.
Connector (JP20): shutter motor
JP36
Connector (JP36): shutter close detection sensor
Connector (JP35): shutter open detection sensor
Grounding cable: 1 hex/Phillips-head screw (M5 x 10)
JP35
Caution
in
"5.2.3 Removing/Installing the Exterior Panel
Chapter 5 Disassembly and Assembly
Always attach the grounding cable in the
direction shown in the figure.
Step
5-39

Hide quick links: