Konica Minolta REGIUS 110 Service Manual page 191

Direct digitizer
Table of Contents

Advertisement

A
First Front Back Side
Chapter 5 Disassembly and Assembly
15
Move the holding plate to the second front side.
16
Remove the encoder unit assembly.
2 hex/Phillips-head screws (M3 x 12)
Move the holding plate on the subscan unit until the
setscrews fixing the encoder bracket reaches the
position A in the figure in order to remove the screws
one by one.
17
Remove the encoder.
3 screws (M3 x 12)
18
Install the new encoder.
3 screws (M3 x 12)
19
See
Step 16
to install the encoder unit assembly.
2 hex/Phillips-head screws (M3 x 12)
Fix the encoder unit assembly by pressing in the
direction of the arrow in the figure.
It will be easier to install if you temporarily
attach the encoder unit assembly using
Note
screws, and then pressing it in the direction
of the arrow in the figure.
20
See
"5.5.2 Replacing the Simple Fixing Unit
(Page 5-67)"
to install the simple fixing unit
assembly.
21
Pull the wire into the equipment carefully so as
not to fold it.
Stick one front edge the wire to the cassette absorption
plate magnet from the second front side, then pull out
the stuck the front edge from the second front back
side.
22
Fix one front edge the wire using the wire presser
plate in the second front back side.
1 screw (M4 x 8)
Install the wire with the level surface in the installation
part down. Install other wire in this way for the second
front side.
5-71