Konica Minolta REGIUS 110 Service Manual page 111

Direct digitizer
Table of Contents

Advertisement

Step Operation Sequence
Step operation sequence will differ depending on the use/no use of the cassette. (Perform the operation
with a
)
Sequence
SeqID
type*
A
B
50
Detection of the cassette
51
Reading the barcode
52
Shutter open &wait for cassette to drop
53
Move the receiver to the first falling position
54
Fix the cassette with the justifier guide at the first falling position
55
Slightly release the cassette fixation by the justifier guide
56
Move the receiver to the second falling position
57
Fix the cassette with the justifier guide at the second falling position
58
Move the push plate unit to the press down position & shutter close
59
Release the lock
Chapter 4 Confirming Operation Using Service Tool
3
Click the [Initialization] button.
Initialization action will be performed, and [Next] button
will be enabled one it is completed.
4
Click the [
] button.
Next
Step operation is performed.
When using the cassette
Click the [Next] button after setting the cassette in the
insertion slot.
Operation check is performed for each step operation,
and the step operation sequence ID and its result will
be displayed in the [Information] area.
See
"
Step Operation Sequence"
operation and the sequence ID.
As an example, if the cassette detection operation was
OK, it will display [Next [50] ... OK].
5
Click the [
] button again to move to next step.
Next
Ejecting a Cassette
If you are using a cassette, the cassette will be ejected
from the insertion slot in sequence ID 71. To move to
next step, click [Next] button after removing the
cassette.
Once the last step (sequence ID 73) is completed,
"Next [All]... OK" is displayed in the [Information] field,
and [Initialization] button becomes available.
Operation
for the step
4-9