Chapter 5 Disassembly and Assembly
6
Disconnect the equipment and the REGIUS console from the network over the site to establish
the peer-to-peer based connection.
7
Turn both the REGIUS console and equipment powers on to get them started.
8
In the "Network Configuration" screen of the service tool, re-set the settings to the values when
network configuration was set up on the equipment (during operations at the site).
9
Restore the settings in the REGIUS console that were backed up in
10
Connect the REGIUS console and the equipment to the network over the site.
11
Make sure that all data was reset to the settings that were made during operations, excluding
each compensation and Config data.
12
In the "Change Circuit Board" screen of the service tool, restore the data backed up in
(or during installation/servicing) in the CF card.
No Data to Restore
If data in the CF card cannot be backed up and there is no data backed up during installation/servicing,
perform the rechecking/readjustment procedures as described below:
Checking the number of read pixels and the read position
Adjusting the image size and position
Unevenness calibration
Sensitivity calibration
For performing each work, see the Installation/Service Manual for REGIUS console.
After configuration is completed, perform backup of the CF card data.
Important
13
Restart the equipment.
Now, you have finished with the procedures to replace the CF card.
5-120