Konica Minolta REGIUS 110 Service Manual page 87

Direct digitizer
Table of Contents

Advertisement

Phenomenon Occurring by Setup Problem
Following phenomenon will occur when there is incorrect setup (like same IP address) in the REGIUS
MODEL 110, REGIUS console, and JM.
Restart after setting up so the IP addresses and host names do not duplicate.
Setup condition
There are 2 REGIUS MODEL 110 with same
IP address and host name.
There are 2 REGIUS MODEL 110 with same
IP address.
There are 2 REGIUS consoles with same host
name.
There are 2 JM with same IP address and host
name.
There are 2 REGIUS consoles with same IP
address.
There are 2 JMs with same IP address.
There are 2 REGIUS MODEL 110s with same
host name.
There are 2 JMs with same host name.
There are REGIUS MODEL 110 and REGIUS
console with same IP address and host name.
There are REGIUS MODEL 110 and JM with
same IP address and host name.
Chapter 3 Troubleshooting
Phenomenon
Both will startup normally from power ON to
"READY" status.
When the second REGIUS MODEL 11 starts the
read action after first REGIUS MODEL 110
performed the read action, error (04214) will occur
when the cassette is inserted, and error (04214)
will occur after that on the first equipment too.
When the cassette is inserted simultaneously on
both equipment, error (13031) will occur.
Read action on the first equipment will be normal, but
when the read action starts on the second equipment,
action will stop after displaying "BUSY" on the
message display window and have error (04202).
Warning saying there are duplicate host name is
displayed in the Windows, and REGIUS console will
startup, but cannot read the image. REGIUS MODEL
110 will operate normally.
It will operate with the JM that has started up first.
Warning saying there are duplicate host name will be
displayed in the Windows, but the read action is
possible. It will take longer to insert the cassette after
detecting the barcode only on the first read.
Attempt to operate from the JM that started up later
will cause a registration error.
REGIUS console that changed the IP address will not
startup. Warning will be displayed in the Windows.
It will operate with the JM that has started up first. It
will take longer to insert the cassette after detecting
the barcode only on the first read.
There is no problem, and it will read normally.
It will operate with the JM that has started up first.
Warning will be displayed in the Windows saying there
are duplicate host name.
If you change the IP address and host name of the
REGIUS MODEL 110 to the one used in the
REGIUS console, REGIUS MODEL 110 will
startup, but it will not read.Error (04214) occurs.
When you change the IP address and host name
of the REGIUS console to the one used in the
REGIUS MODEL 110, REGIUS console will
startup, but there will be error (04214) on the
REGIUS MODEL 110 when the cassette is
inserted, and application error will occur on the
REGIUS console too.
When you change the IP address and host for the
REGIUS MODEL 110 to the one used in the JM,
error (24000) will occur.
When you change the IP address and host name
of the JM to the one used in the REGIUS MODEL
110, both REGIUS MODEL 110 and application
will not startup.
3-33