Konica Minolta REGIUS 110 Service Manual page 208

Direct digitizer
Table of Contents

Advertisement

Chapter 5 Disassembly and Assembly
15
See
"6.5 Adjust Position of Detach Detection Roller/Sensor (Page 6-12)"
and sensor positions.
16
See
"
Installation Procedures (Page 5-14)"
install the optical unit.
Now, you have finished with the procedures to replace the detach detection sensor (lower).
5-88
7
Remove the roller retention lower unit assembly.
2 screws (M3 x 6)
To access the screws, slide the sensor mount board
upward.
8
Disconnect the spring connected to the sensor
mount board, from one end of the spring.
9
Remove the sensor mount board.
2 screws (M3 x 6)
1 guide spacer for each
10
Remove the detach detection sensor (lower).
1 screw (M3 x 6)
11
Install the new detach detection sensor (lower).
12
See
Step 9
and
Step 8
board.
2 screws (M3 x 6)
1 guide spacer for each
Spring
The setscrew removed in
when the sensor position will be adjusted
13
Install the roller retention unit lower assembly that
was removed in
Step
2 screws (M3 x 6)
14
Install the detach detection roller that was
removed in
Step
3.
in
"5.2.5 Removing/Installing the Optical Unit"
to install the sensor mount
Step 6
will be installed later
(Step
15).
7.
to adjust the roller
to