Chapter 5 Disassembly and Assembly
Second Front Side
5-70
ACN4
11
Loosen the setscrew on the wire presser plate in
the second front back side to remove the wire.
1 screw (M4 x 8)
Loosen the screw.
Put down the removed wire without pulling at it or
anything.
Handle the wire carefully so as not to
damage or fold it.
Caution
12
Loosen the setscrew on the wire presser plate in
the second from side to remove the wire.
1 screw (M4 x 8)
Loosen the screw.
13
Disconnect the wire from the encoder.
If the encoder is not replaced, skip to
If the encoder is replaced, stick the removed wire to
the cassette absorption plate, making sure the wire is
not folded back.
14
Cut off the wiring band binding the encoder cable
(connector: ACN4) and the other cables.
The wired optical unit is shown in the figure for
reference, though the optical unit has been already
removed.
Step
21.