Konica Minolta REGIUS 110 Service Manual page 215

Direct digitizer
Table of Contents

Advertisement

JP1
Second Front
First Front Side
B
B
A
B
Second Front Back Side
White
Black
Black White
4
Remove the connector (JP1) from the first front
side.
Caution
5
Remove the grounding cable from the second
front side.
1 hex/Phillips-head screw (M5 x 10)
Back Side
6
Remove the exterior frame (right).
A: 3 hex/Phillips-head screws (M4 x 8)
B
B: 14 screws (M4 x 8)
Remove all screws.
B
7
Remove the noise filter and the circuit protector
unit assembly.
Noise filter
2 screws (M4 x 8)
Grounding cable: 1 screw (M4 x 8)
Circuit protector unit assembly
2 screws (M4 x 8)
8
Replace the parts.
Circuit breaker cover
1 screw (M4 x 8)
Inlet
2 screws (M3 x 6)
Install the circuit protector (power supply circuit
Blue
breaker) with the ON side up.
Yellow
Install the inlet with the ground side facing to the left.
Chapter 5 Disassembly and Assembly
When removing the connector, hold a notch
on the connector without pulling a cable.
Giving a strong pull to the cable may cause
it to be broken.
5-95

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

0902

Table of Contents