Konica Minolta REGIUS 110 Service Manual page 216

Direct digitizer
Table of Contents

Advertisement

Chapter 5 Disassembly and Assembly
15
See
"
Installation Procedures (Page 5-8)"
and Insertion Unit"
Now, you have finished with the procedures to replace the noise filter and the circuit protector unit.
5-96
9
See
assembly.
Move the noise filter in the direction of the arrow in the
figure and fix it.
10
See
11
See
inlet to the frame.
12
Connect the connector (JP1) that was removed in
Step
13
Connect the CIU connector (CN5) that was
removed in
14
Connect the cooling fan connector (JJ62) that
was removed in
in
to install the insertion unit and exterior panel.
Step 7
to install the locking pin unit
Noise filter
2 screws (M4 x 8)
Grounding cable: 1 screw (M4 x 8)
Circuit protector unit assembly
2 screws (M4 x 8)
After installing, make sure that circuit
protector has been switched to OFF.
Caution
Step 6
to install the exterior frame (right).
A: 4 hex/Phillips-head screws (M4 x 8)
B: 12 screws (M4 x 8)
Step 5
to connect the grounding cable of the
1 hex/Phillips-head screw (M5 x 10)
4.
Step
3.
CN5 (CCN5: cable)
Step
2.
"5.2.3 Removing/Installing the Exterior Panel