Konica Minolta REGIUS 110 Service Manual page 150

Direct digitizer
Table of Contents

Advertisement

Chapter 5 Disassembly and Assembly
13
See
"
Installation Procedures (Page 5-8)"
and Insertion Unit"
Now, you have finished with the procedures to replace the insertion slot detection sensor (LPB).
5-30
Grounding Cable
JP36
JP35
JP20
to install the insertion unit and exterior panel.
10
Install the shutter unit assembly that was
removed in
Step
3.
5 tapping screws (M4 x 10)
Press it in a direction of the arrow shown in the figure
to fix it.
11
Use a 0.2 mm-spacer (film) to check the clearance
between the lock cam and the stopper block.
If the clearance is any other value than 0.2 mm, adjust
the position of the stopper block.
12
Connect the cables that were unplugged in
2
to the shutter unit assembly.
Connector (JP20): shutter motor
Connector (JP35): shutter close detection sensor
Connector (JP36): shutter open detection sensor
Grounding cable: 1 hex/Phillips-head screw (M5 x 10)
Always attach the grounding cable in the
direction shown in the figure.
Caution
in
"5.2.3 Removing/Installing the Exterior Panel
Step

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

0902