Table of Contents

Advertisement

SPECTROPHOTOMETER
CM - 3700A
INSTRUCTION MANUAL
Before using this instrument,
please read this manual.

Advertisement

Table of Contents
loading

Summary of Contents for Konica Minolta CM-3700A

  • Page 1 SPECTROPHOTOMETER CM - 3700A INSTRUCTION MANUAL Before using this instrument, please read this manual.
  • Page 2: Safety Symbols

    Microsoft Corporation in the United States and other countries. Notes on this Manual • Copying or reproduction of all or part of the contents of this manual without KONICA MINOLTA’s permis- sion is strictly prohibited.
  • Page 3: Safety Precautions

    Should liquid or metal objects AC adapter other than those specified by enter the instrument, turn the power KONICA MINOLTA is used, it may result in OFF immediately, disconnect the damage to the unit, fire or electric shock. AC adapter plug from the AC outlet, and contact the nearest Konica Minolta-authorized service facility.
  • Page 4: Introduction

    Keep all packing materials (cardboard box, cushioning material, plastic bags, etc.) in a safe place. The CM-3700A is a precision measuring instrument. They can be used to protect the instrument from impact and vibration during shipment to Konica Minolta for maintenance.
  • Page 5: Notes On Use

    Power Source • Make sure that the power switch is set to OFF (“ I “) when the CM-3700A is not in use. • Always use the AC adapter supplied as a standard accessory (AC-A308) and connect it to an AC outlet of the rated voltage and frequency.
  • Page 6 • The specimen measuring port and integrating sphere are extremely precise components, and great care should be taken to prevent them getting dirty or exposing them to impacts. When the CM-3700A is not in use, be sure to attach a target mask to the measuring port to prevent entry of foreign matter.
  • Page 7: Notes On Storage

    And store them in the accessory case (CM-A156). • Take care not to leave the CM-3700A for a long period of time with a target mask attached. The sample holder may stick to the target mask.
  • Page 8: Table Of Contents

    Measurement Procedure ....... 12 About haze measurements ......... 26 Flow of Preparation and Measurement ....12 Cleaning the CM-3700A and Accessories ...27 Connecting a Personal Computer ..13 Zero Calibration Box and White Calibration Plate.. 27 Target Mask ............27 Connecting the AC Adapter ....
  • Page 9: Using The Cm-3700A

    Using the CM-3700A...
  • Page 10: Standard Accessories

    Used to perform zero calibration for measurement of reflectance and to perform haze measurement at measurement of transmittance. AC Adapter AC-A308 Used to supply power from an AC outlet to the CM-3700A. Input: 100 to 240 V 50/60 Hz Output: 8 V 1.5 A Plug design:...
  • Page 11: Optional Accessories

    SpectraMagic™ NX CM-S100w This software provides various functions (e.g., data processing and file management) and allows the user to operate the CM-3700A using a personal computer. Transmittance Specimen Holder CM-A96 Used to hold the specimen for measurement of transmittance. It can hold specimens up to 22.5 mm (7/8 in.) thick.
  • Page 12: System Diagram

    USB Cables CM-A91 (SAV) IF-A21 ( 3m) IF-A22 ( 5m) (3×5 mm measurement / 5×7 mm illumination) Spectrophotometer Target Mask  CM-3700A CM-A92 (MAV) (ø8 mm measurement / ø11 mm illumination) AC Adapter AC-A308 Target Mask  CM-A93(LAV) (ø25.4 mm measurement / ø30 mm illumination))...
  • Page 13: Names And Functions Of Parts

    ⑦ Transmittance specimen chamber cover ..Covers the transmittance specimen chamber when slid closed. ⑧ USB connection terminal ......Used to connect the instrument to a PC with the supplied USB cable (IF-A21 or IF-A22). ⑨ AC adapter input socket......Used to connect the AC adapter supplied with the CM-3700A.
  • Page 14: Measurement Procedure

    Measurement Procedure c This manual explains how to prepare the CM-3700A and how to set a specimen. c The CM-3700A is controlled by a PC to perform measurements. c For a description of measuring method using SpectraMagic™ NX (optional), refer to the SpectraMagic™ NX instruction manual.
  • Page 15: Connecting A Personal Computer

    • The USB communication port of the instrument conforms to USB 1.1. { To connect the instrument to a PC, you need to install the USB driver dedicated to the CM-3700A. Install the USB driver supplied with the software that enables connection and operation of the instrument.
  • Page 16: Connecting The Ac Adapter

    Failure to do so may damage the CM- 3700A or the AC adapter, causing a fire or electric shock. If the CM-3700A will not be used for a long time, disconnect the AC adapter from the AC outlet. Accumulated dirt or water on the prongs of the AC adapter's plug may cause a fire and should be removed.
  • Page 17: Turning Power On And Off

    Turning Power ON and OFF WARNING The CM-3700A should not be operated if the CM-3700A or the AC adapter is damaged, or smoke or strange odors occur. Doing so may result in a fire. In such situations, turn the power OFF immediately, disconnect the AC adapter from the AC outlet, and contact the nearest Konica Minolta- authorized service facility.
  • Page 18: Attaching A Target Mask

    { When not using the CM-3700A, attach one of the target masks or the protective cap to prevent dust entering the integrating sphere. { Take care not to leave the CM-3700A for a long period of time with a target mask attached. The sample holder may stick to the target mask.
  • Page 19: Calibration Of Reflectance And Transmittance Measurements

    { The EEPROM of the Spectrophotometer has space for one set of calibration data for transmittance measurements.  100% calibration data Because it is necessary to set the 100% transmittance data to the CM-3700A for all wavelengths, 100% calibration data are normally set to all wavelengths at shipment.
  • Page 20: Performing Calibration

     Performing calibration Calibration (both zero calibration and white calibration or both 0% calibration and 100% calibration) is effective only while the Spectrophotometer remains switched on. If it is switched off and then switched on again, both zero calibration and white calibration (or both 0% calibration and 100% calibration) must be performed before measurements are taken. c The calibration becomes effective only when being performed under six kinds of measurement conditions.
  • Page 21: Attaching The Zero Calibration Box

    Do not perform measurement with the specimen measuring port directed towards your face. Doing so may cause damage to your eyes . Be careful around openings in the CM-3700A. Failure to do so may result in fingers being trapped causing injury.
  • Page 22: Attaching The White Calibration Plate

    Do not perform measurement with the specimen measuring port directed towards your face. Doing so may cause damage to your eyes . Be careful around openings in the CM-3700A. Failure to do so may result in fingers being trapped causing injury.
  • Page 23: Attaching The Transmittance Zero Calibration Plate (Optional Accessory)

    Attaching the Transmittance Zero Calibration Plate (Optional Accessory) The Transmittance Zero Calibration Plate is used for performing zero calibration for transmittance measurements. c Before starting transmittance calibration and measurements, attach the LAV (ø25.4 mm) target mask and position the white calibration plate at the reflectance measuring port. c Before starting transmittance calibration and measurements, set the measurement area to LAV and the specular component to SCI using the software.
  • Page 24: Performing 100% Calibration

    Performing 100% calibration Two methods for performing 100% calibration for transmittance measurements are possible: • 100% calibration to air: When the specimen to be measured is in sheet or solid form, 100% calibration should be performed with the transmittance chamber empty. •...
  • Page 25: Setting A Specimen

    Doing so may cause a fire. Do not disassemble or modify the CM-3700A. Doing so may cause a fire or electric shock. The CM-3700A should not be operated if it is damaged, or smoke or strange odors occur. Doing so may result in a fire.
  • Page 26: The Sample Holder For Measuring The Reflectance

    When widely opening the measurement port baseplate { If you remove the stopper below the measurement port baseplate, the measurement port baseplate can two screws be widely opened. { Remove the two stopper attaching screws by turning them counterclockwise using a Phillips-head screwdriver. { Keep the removed screws and stopper in a safe place.
  • Page 27: Transmittance Measurements

    { Take care not to spill the liquid specimen on the CM-3700A. Should there be any spillage, immediately wipe it with a soft, dry cloth.
  • Page 28: The Transmittance Chamber For Transmittance Measurement

     The transmittance chamber for transmittance measurement Set the liquid and transmittance specimen to the transmittance chamber. If you cannot set the specimen because it touches the Light Trap in the transmittance chamber, remove the Light Trap Transmittance chamber before setting the specimen. Target Mask (LAV) and White Calibration Plate...
  • Page 29: Cleaning The Cm-3700A And Accessories

    Doing so may cause a fire or electric shock. The CM-3700A should not be operated if it is damaged, or if smoke or strange odors occur. Doing so may result in a fire. In such situations, turn the power OFF immediately, disconnect the AC adapter from the AC outlet, and contact the nearest Konica Minolta-authorized service facility.
  • Page 30: Receiving Window

     Receiving Window Set the measurement area to SAV using your software. Memo • This will cause the lens to move closer to the receiving window, making it easier to clean. Use a blower to remove dirt and dust from the Receiving window receiving window.
  • Page 31: Error Message

    NX CM-S100w, optional color ® management software, to control the instrument. If you see an error message, follow the instructions shown in the table below. If a problem persists, contact the nearest Konica Minolta-authorized service facility. Error Messege Problem / possible cause Action...
  • Page 32 No response from Communication with Turn on the power of the instrument the instrument failed. instrument. • The power of the Check and correct instrument is off. the COM port and • COM port setting communication settings. is incorrect. • Communication settings are incorrect.
  • Page 33: Troubleshooting Guide

    TROUBLESHOOTING GUIDE If a problem occurs with the Spectrophotometer, please check the following points before requesting service. If the problem continues to occur even after the suggested corrective actions have been taken, contact the nearest Konica Minolta-authorized service facility. Refer Condition...
  • Page 35: Explanations

    Explanations...
  • Page 36: Liiumination/Observation System

    The flow of measurement is shown below. The geometry of the CM-3700A conforms to CIE No.15, ISO 7724/1, ASTM E 1164, DIN 5033 Teil 7, and JIS Z 8722 condition c (diffused illumination/perpendicular viewing system) standards, and offers both di:8° (SCI: specular component included ;...
  • Page 37: Illumination And Measurement Areas

    Illumination and Measurement Areas The CM-3700A allows you to select a target mask from three types: LAV (for ø25.4mm measurements), MAV (for ø8mm measurements) and SAV (for 3×5mm measurements), according to the specimen and your application. Select and attach a suitable target mask (illumination area) for each measurement area.
  • Page 38: Fluorescent Measurement

    Fluorescent Measurement The Spectrophotometer is equipped with a UV cutoff filter positioned in front of the pulsed xenon arc lamp. The position of this filter is controlled by a motor and allows the amount of UV included in the illumination to be controlled.
  • Page 39: Dimensions

    Dimensions (mm) 71.8 91.5...
  • Page 40: Specifications

    Max E* 0.3 (corresponds to approx. E* 0.2) for any of 12 BCRA Series II agreement color tiles compared to values measured with Konica Minolta master body. UV adjustment Computer controlled: continuously variable Maximum sample thickness: Approx. 50 mm Transmittance...
  • Page 41 < CAUTION > KONICA MINOLTA WILL NOT BE LIABLE FOR ANY DAMAGES RESULTING FROM THE MISUSE, MISHANDLING, UNAUTHORIZED MODIFICATION, ETC. OF THIS PRODUCT, OR FOR ANY INDIRECT OR INCIDENTAL DAMAGES (INCLUDING BUT NOT LIMITED TO LOSS OF BUSINESS PROFITS, INTERRUPTION OF BUSINESS, ETC.) DUE TO THE...
  • Page 42 © 2011-2017 KONICA MINOLTA, INC. 9222-A4KX-11 BHEEDK...

Table of Contents