Section 4: Service menu and Factory Default Settings

4.1 Introduction

4.2 Service Menu

4.3 Demo Mode
4.4 Factory Default Settings
4.1 Introduction
This section discusses use of the Service menu to configure Power-on default settings,
Alarm Suspend, Alarm Silence Period, Audible Alarm Type, AC Line Frequency,
Language selections, NIBP Test Mode access, and System Information to obtain
service-related information about the monitor. Also this section explains briefly the
factory default settings and Demo mode.
4.2 Service Menu
The purpose of the Service menu (Figure 6, Table 9) is to allow the authorized user to
create a Power-on default for each setting in the N5500. Power on defaults are the
settings in effect each time the N5500 is powered on. Once the Service menu is entered,
physiological monitoring is terminated. The screen layouts do not display any
information associated with normal monitoring operation. Use the following procedure
to configure the Service Menu for the N5500 monitor (also see Using the Monitor
section, page 35 of the operator' s manual):
1. Set N5500 to normal monitoring mode.
2. Rotate the knob to highlight the Set-up icon. (Set-up icon located on the bottom of
ECG
1mV
SET UP MENU
Patinet Mode : Adult
Trend Clear
Print Speed : 25.0 mm/s
Print Mode : One-Shot
ECG Cable Type : 3 Leads
Alarm Volume : 5
Key Beep Volume : 4
QRS Volume : 4
Service Menu
Return
Adult
1. Set-up icon
Figure 5. The access of Service Menu via Set-up menu
N5500 Service Manual
HR
120
ECG
SpO
2
100
0
NIBP
0
0
140/100
(120)
RR
Return
37.0
T
1
C
T
2
F
06/12/2002 08:12:23
180
24
29