F a c t o r y D e f a u l t s
Neonate Default Settings
N-595
The N-595 is shipped with factory default settings. Authorized
technical personnel using the procedures described in the N-595
service manual can change default settings.
Table 11: Neonate Factory Defaults
Parameter
Monitoring Mode
%SpO
Lower Alarm Limit
2
%SpO
Upper Alarm Limit
2
Allow silence duration to be set to
OFF
Alarm Silence Duration
Alarm Silence Reminder
Alarm Volume
Data Port Baud Rate
Data Port Protocol
Display Contrast
Display Format
Language
Nurse Call Polarity
Pulse Beep Volume
Pulse Rate Lower Alarm Limit
Setting
Neo
80%
95%
Yes
60 seconds
Enabled
7 of 10
9600
ASCII
Midrange
Pleth
English
Normally Low
4 of 10
90 beats per minute
139