Battery Replacement

Removal

N-595
1. Follow the procedure in paragraphs Prior to Disassembly on page 64, and
Monitor Disassembly on page 65.
2. Remove the two screws from the battery bracket and lift the battery out of the
bottom case as shown in Figure 7.
3. Be sure to note the polarity of the leads. Use needle-nose pliers to disconnect
the leads from the battery.

Figure 7: Removing the Battery

4. The lead-acid battery is recyclable. Do not dispose of the battery by placing it
in the regular trash. Dispose of the battery in accordance with local guidelines
or return it to Nellcor's Technical Services for disposal.