Disassembling the Monitor

Assembling the Monitor

Battery Replacement

Removing the Old Battery

88
To assemble the monitor:
1. Connect the monitor's Power Supply to J200 on the Main Board PCB.
2. Place the monitor's top enclosure over the bottom case, being careful to align
the Display PCB, Power Entry Module, and the fan with the slots in the
monitor casing.
Caution: Ensure not to pinch any wires between the cases when reassembling
the monitor. Inspect the entire edge of the union for even seating before
installing the screws.
3. Place the screws in the bottom cover of the monitor and gently tighten the
screws that hold the case halves together (10-in lbs. maximum).
Caution: Overtightening could strip out the screw holes in the top enclosure,
rendering it unusable.
4. Replace the four corner screws.
To remove the old battery:
1. Turn the monitor off by pressing the ON/STANDBY button.
2. Disconnect the monitor from the AC power source.
3. Follow the steps in Disassembling the Monitor on page 85.
4. Remove the two screws from the battery bracket and lift the battery out of the
bottom case as shown in Figure 7 on page 89.