9 Operation at Use of a PC
9.3 Configuration of the Device
Table 9-15
Total recording duration
Pretrigger ratio
Record ph-ph voltage
1)
The Detection mode will always synchronize with the setting of Record ph-ph voltage under Waveform
capture setting.
Parameterizing the Trigger Management
To change the trigger management settings, proceed as follows:
In the navigation window, select the Recording and Reporting menu and click Trigger management.
Figure 9-47
Select the respective parameters in the list boxes according to Table 9-15.
After the parameterization click Send.
The parameters are transmitted to the device but not enabled yet (passive set of parameters).
If you do not want to make any additional settings, continue with the Activation of the device configuration
according to section Activating the Set of Parameters. If you want to change other settings, enter the
changes and then enable the device configuration as described in section Activating the Set of
Parameters.
196
Trigger Management
2.0 s
10 %
no
Configure Tab, Trigger Management Input/Output Window
0.2 s to 3.0 s in 0.2-s steps
0 % to 30 % in 5-% steps
no
yes
SICAM Q100, 7KG95xx, Device Manual
E50417-H1040-C522-A6, Edition 06.2019