Time Base Accuracy Test

Specification

Electrical Characteristics
Frequency accuracy

Description

This test measures the accuracy of the power meter's time base frequency.
Time base frequency determines the accuracy of the sampling rate when
making measurements. Time base frequency is measured by attaching a
frequency counter to the Trig Out socket on the rear of the power meter then
enabling test mode from the service menu.

Equipment

• Frequency counter: Agilent 53132A.
E4416A/E4417A Power Meters Service Guide
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Performance Tests
Performance Limits
10 MHz ±0.01%