6.3.4

Measuring Rate

The maximum sampling rate is 15 kHz.
The selection of the measuring rate is made in the menu
measuring rate is continuously adjustable within a range from 0.1 kHz to 5 kHz. The increment is 1 Hz.
Changing the Measuring Rate
Switch to the tab
Settings
Select the desired measuring rate.
Observing the video signal is useful for selecting the measuring rate.
If the video signal or the setup signal is output via an interface, the measuring rate is reduced to 100 Hz.
6.3.5

Frame Averaging

Frame averaging is activated by default (= ON). It provides an optimum signal-noise ratio.
ON: Several video recordings are created, averaged, and undergo further processing.
OFF: One video recording is created and undergoes further processing. This is necessary for fast-moving targets.
Deactivating Frame Averaging
Switch to the tab
Settings
For fast-moving objects, switch frame averaging off (= OFF).
When operating without the frame averaging function, the sensor automatically varies the exposure time for the line scan
camera.
6.3.6

Counter Reset

You can reset the measured value counter or the timestamp by pressing the relevant button.
Counter Reset Procedure
Switch to the menu
optoCONTROL 2700
>
Data acquisition
>
Data acquisition
>
Settings
Data acquisition
>
Settings
Data recording
>
Measuring
rate.
>
Frame
averaging.
>
and click on the relevant button.
Counter reset
Advanced Settings
>
rate. The
Measuring
Seite 48